Fast Beam Delivery for Flash Irradiations at the HZB Cyclotron |
178 |
|
- J. Bundesmann, A. Denker, G. Kourkafas
HZB, Berlin, Germany
- J. Heufelder, A. Weber
Charite, Berlin, Germany
- P. Mühldorfer
BHT, Berlin, Germany
|
|
|
In the context of radiotherapy, Flash irradiations mean the delivery of high dose rates of more than 40 Gy/s, in a short time of less than one second. The expectation of the radio-oncologists are lesser side effects while maintaining the tumour control when using Flash. Clinically acceptable deviations of the applied dose to the described dose are less than 3%. Our accelerator control system is well suited for the standard treatment of ocular melanomas with irradiaton times of 30 s to 60 s. However, it is too slow for the short times required in Flash. Thus, a dedicated beam delivery control system has been developed, permitting irradiation times down to 7 ms with a maximal dose variation of less than 3%.
|
|